Maintenance and qualification of analytical instruments

Preventive and corrective maintenance of chromatography systems, and IQ/OQ/PQ qualification of HPLC systems from any manufacturer.

Preventive maintenance is essential to keep chromatography systems performing at their best in an environment that complies with good laboratory practice. It verifies that the instrument operates within the manufacturer's parameters and leaves documented evidence of its performance.

What maintenance includes

  1. 1Replacement of pump and injector seals.
  2. 2Internal and external cleaning of every module in the system.
  3. 3Detector lamp replacement, when needed.
  4. 4Diagnostic tests to confirm the system works correctly after service.

Benefits

  • Lower costs

    More affordable service, with flexible maintenance contracts that fit the laboratory's budget.

  • Flexible scheduling

    We adapt to your schedule and qualify outside operating hours to minimize downtime.

  • Multi-brand expertise

    Our engineers know instruments from many manufacturers, which makes integration with the rest of the lab easier.

  • Custom protocols

    We design qualification protocols to match each laboratory's regulations and type of analysis.

  • Faster technical support

    Agile, personal attention with shorter waits for support or maintenance.

  • Lower-cost spare parts

    Compatible alternatives that don't compromise quality, without costly manufacturer contracts.

HPLC qualification phases (any manufacturer)

  1. Phase 1

    Installation qualification (IQ)

    • Verification that the system is installed correctly.
    • Review of components: pump, injector, column, detector and software.
    • Check of electrical and flow connections.
    • Record of serial numbers and models for every module.
  2. Phase 2

    Operational qualification (OQ)

    • Tests confirming the instrument operates within expected values.
    • Pump flow verification (precision and accuracy).
    • System pressure check.
    • Autosampler precision (injected volume).
    • Detector test (linearity and wavelength accuracy).
    • Column oven temperature control, where applicable.
  3. Phase 3

    Performance qualification (PQ)

    • Tests with standards to assess repeatability and accuracy.
    • Injection of a standard (caffeine, benzoic acid or other) compared against reference values.
    • Evaluation of peak resolution, retention time and symmetry.
    • Reproducibility test with multiple injections of the same standard.
  4. Phase 4

    Periodic requalification

    • Scheduled quarterly, semi-annually or annually.
    • Ensures the instrument keeps operating within specification.

Frequently asked questions

How often should preventive maintenance be done?

We recommend every six months or once a year. It depends on how heavily the chromatography system is used and on each laboratory's needs.

Can generic spare parts be used? What warranty do they carry?

Yes. Generic parts are an alternative to the original manufacturer's: they are designed to be comparable in form, fit and function to the parts they replace, without compromising service quality. They carry a one-year warranty from installation.

Which standards and measuring equipment are used for calibration?

Certified standards traceable to NIST, and measuring equipment calibrated by laboratories accredited by EMA (the Mexican accreditation body).

How experienced are the service engineers?

Our team is trained directly by the instrument manufacturers and by institutions such as CENAM. We have supported analytical instruments for 20 years.